Journal of Orthopaedic Research: Biocompatibility, bone healing, and safety evaluation in rabbits with an IlluminOss bone stabilization system

McSweeney AL, Zani BG, Baird R, Stanley JR, Hayward A, Markham PM, Kopia GA, Edelman ER, Rabiner R. Biocompatibility, bone healing, and safety evaluation in rabbits with an IlluminOss bone stabilization system

Summary: Cone healing, biocompatibility, and safety employing the IlluminOss System (IS), comprised of an inflatable balloon filled with photopolymerizable liquid monomer, was evaluated in New Zealand white rabbits. Successful bone healing and callus remodeling over 6 months was demonstrated radiologically and histologically with IS implants in fenestrated femoral cortices. Biocompatibility was demonstrated with IS implants in brushed, flushed femoral intramedullary spaces, eliciting no adverse, local, or systemic responses and with similar biocompatibility to K-wires in contralateral femurs up to 1 year post-implant. Lastly simulated clinical failures demonstrated the safety of IS implants up to 1 year in the presence of liquid or polymerized polymer within the intramedullary space. Polymerized material displayed cortical bone and vasculature effects comparable to mechanical disruption of the endosteum. In the clinically unlikely scenario with no remediation or polymerization, a high dose monomer injection resulted in marked necrosis of cortical bone, as well as associated vasculature, endosteum, and bone marrow. Overall, when polymerized and hardened within bone intramedullary spaces, this light curable monomer system may provide a safe and effective method for fracture stabilization.

J Orthop Res. 2017 Oct;35(10):2181-2190. doi: 10.1002/jor.23532.

October 1, 2017